CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability

[RECALL]

Jun 9, 2026 · CMR Surgical · Versius · India

Failure Mode

RECALL

Event Date

Jun 9, 2026

Primary Entity

CMR Surgical

Robotic Platform

Versius

Jurisdiction / Location

India

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

CMR Surgical initiated a Class 2 recall (Z-2717-2026) of two Versius Surgeon Console units (Model V-SS-1020) because Secure Boot was mistakenly not enabled at manufacturing time, presenting a potential cybersecurity risk. The affected units were distributed in California; the firm began notifying customers on June 16, 2026, and will proactively configure Secure Boot on the affected consoles.

System & Fleet Architecture

Operating Manufacturer

CMR SurgicalView Manufacturer

Hardware Platform

Versius

Operational Sectors

Primary Sources & Evidentiary Filings

  • https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=99280

    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=99280

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?

CMR Surgical initiated a Class 2 recall (Z-2717-2026) of two Versius Surgeon Console units (Model V-SS-1020) because Secure Boot was mistakenly not enabled at manufacturing time, presenting a potential cybersecurity risk. The affected units were distributed in California; the firm began notifying customers on June 16, 2026, and will proactively configure Secure Boot on the affected consoles.

When did this incident occur?

The incident is recorded as occurring on June 9, 2026 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?

Versius by CMR Surgical is the recorded robot involved in this incident at India.

Has anyone responded to CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.