da Vinci Hot Shears (monopolar curved scissors) Class I recall: insulation micro-cracks caused tissue burns

[RECALL]

May 8, 2013 · Intuitive Surgical · da Vinci (and Ion) · South Korea

Failure Mode

RECALL

Event Date

May 8, 2013

Primary Entity

Intuitive Surgical

Robotic Platform

da Vinci (and Ion)

Jurisdiction / Location

South Korea

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

Class I recall (Urgent Medical Device Notification May 8 2013): micro-cracks near the distal shaft on '-09'/'-10' versions of the EndoWrist Hot Shears develop after cleaning/sterilization, creating an arc path for monopolar energy to leak and burn tissue outside the surgical field (134 tip-cover complaints + 82 MDRs, Jan 2010-Dec 2011). Instrument versions corrected/superseded. The load-bearing historical da Vinci safety event.

System & Fleet Architecture

Operating Manufacturer

Intuitive SurgicalView Manufacturer

Hardware Platform

da Vinci (and Ion)

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.medscape.com/viewarticle/804067

    https://www.medscape.com/viewarticle/804067

  • https://www.youhavealawyer.com/blog/2013/05/13/da-vinci-robot-recall/

    https://www.youhavealawyer.com/blog/2013/05/13/da-vinci-robot-recall/

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in da Vinci Hot Shears (monopolar curved scissors) Class I recall: insulation micro-cracks caused tissue burns?

Class I recall (Urgent Medical Device Notification May 8 2013): micro-cracks near the distal shaft on '-09'/'-10' versions of the EndoWrist Hot Shears develop after cleaning/sterilization, creating an arc path for monopolar energy to leak and burn tissue outside the surgical field (134 tip-cover complaints + 82 MDRs, Jan 2010-Dec 2011). Instrument versions corrected/superseded. The load-bearing historical da Vinci safety event.

When did this incident occur?

The incident is recorded as occurring on May 8, 2013 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in da Vinci Hot Shears (monopolar curved scissors) Class I recall: insulation micro-cracks caused tissue burns?

da Vinci (and Ion) by Intuitive Surgical is the recorded robot involved in this incident at South Korea.

Has anyone responded to da Vinci Hot Shears (monopolar curved scissors) Class I recall: insulation micro-cracks caused tissue burns?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of da Vinci Hot Shears (monopolar curved scissors) Class I recall: insulation micro-cracks caused tissue burns?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.