da Vinci 8mm SureForm 30 Gray stapler reload recall after incomplete staple lines (1 death, 4 serious injuries)

[RECALL]

Mar 11, 2026 · Intuitive Surgical · da Vinci (and Ion) · South Korea

Failure Mode

RECALL

Event Date

Mar 11, 2026

Primary Entity

Intuitive Surgical

Robotic Platform

da Vinci (and Ion)

Jurisdiction / Location

South Korea

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

Intuitive recalled 8mm SureForm 30 Gray reloads (2mm staples for extra-thin tissue) after incomplete staple lines on blood vessels despite a 'completed' fire indication, linked to one patient death and four serious injuries (bleeding). Customer letter / quarantine Mar 11 2026; FDA classified Class I (reported May 2026).

System & Fleet Architecture

Operating Manufacturer

Intuitive SurgicalView Manufacturer

Hardware Platform

da Vinci (and Ion)

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.medtechdive.com/news/intuitive-surgical-recalls-stapler-reloads-over-issue-linked-to-patient-dea/815158/

    https://www.medtechdive.com/news/intuitive-surgical-recalls-stapler-reloads-over-issue-linked-to-patient-dea/815158/

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in da Vinci 8mm SureForm 30 Gray stapler reload recall after incomplete staple lines (1 death, 4 serious injuries)?

Intuitive recalled 8mm SureForm 30 Gray reloads (2mm staples for extra-thin tissue) after incomplete staple lines on blood vessels despite a 'completed' fire indication, linked to one patient death and four serious injuries (bleeding). Customer letter / quarantine Mar 11 2026; FDA classified Class I (reported May 2026). Blue/white reloads unaffected. Root cause stated 'still under investigation.' Cap-flag: exact recall Z-number + unit count not captured from primary FDA pages. Distinct from the 2022 SureForm fragment recall.

When did this incident occur?

The incident is recorded as occurring on March 11, 2026 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in da Vinci 8mm SureForm 30 Gray stapler reload recall after incomplete staple lines (1 death, 4 serious injuries)?

da Vinci (and Ion) by Intuitive Surgical is the recorded robot involved in this incident at South Korea.

Has anyone responded to da Vinci 8mm SureForm 30 Gray stapler reload recall after incomplete staple lines (1 death, 4 serious injuries)?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of da Vinci 8mm SureForm 30 Gray stapler reload recall after incomplete staple lines (1 death, 4 serious injuries)?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.