Intuitive Surgical recalls da Vinci X and Xi Universal Surgical Manipulators over cannula mount screws that can break (FDA Z-2046-2026)

[RECALL]

Apr 2, 2026 · Intuitive Surgical · da Vinci (and Ion) · South Korea

Failure Mode

RECALL

Event Date

Apr 2, 2026

Primary Entity

Intuitive Surgical

Robotic Platform

da Vinci (and Ion)

Jurisdiction / Location

South Korea

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

On April 2, 2026, Intuitive Surgical initiated a Class 2 recall (FDA Z-2046-2026) of da Vinci X and Xi Surgical System Universal Surgical Manipulators (USMs) because screws within the arm sub-assembly may be susceptible to breaking. If all six screws break, the cannula mount can separate from the USM, potentially causing imprecise or unintended instrument tip motion during surgery. Intuitive sent Urgent Medical Device Correction letters to customers.

System & Fleet Architecture

Operating Manufacturer

Intuitive SurgicalView Manufacturer

Hardware Platform

da Vinci (and Ion)

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?id=219087

    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?id=219087

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in Intuitive Surgical recalls da Vinci X and Xi Universal Surgical Manipulators over cannula mount screws that can break (FDA Z-2046-2026)?

On April 2, 2026, Intuitive Surgical initiated a Class 2 recall (FDA Z-2046-2026) of da Vinci X and Xi Surgical System Universal Surgical Manipulators (USMs) because screws within the arm sub-assembly may be susceptible to breaking. If all six screws break, the cannula mount can separate from the USM, potentially causing imprecise or unintended instrument tip motion during surgery. Intuitive sent Urgent Medical Device Correction letters to customers.

When did this incident occur?

The incident is recorded as occurring on April 2, 2026 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in Intuitive Surgical recalls da Vinci X and Xi Universal Surgical Manipulators over cannula mount screws that can break (FDA Z-2046-2026)?

da Vinci (and Ion) by Intuitive Surgical is the recorded robot involved in this incident at South Korea.

Has anyone responded to Intuitive Surgical recalls da Vinci X and Xi Universal Surgical Manipulators over cannula mount screws that can break (FDA Z-2046-2026)?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of Intuitive Surgical recalls da Vinci X and Xi Universal Surgical Manipulators over cannula mount screws that can break (FDA Z-2046-2026)?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.