Intuitive Surgical recalls 144 da Vinci 5 systems over surgeon console motor connector failure

[RECALL]

Aug 21, 2024 · Intuitive Surgical · da Vinci (and Ion) · Hong Kong, China

Failure Mode

RECALL

Event Date

Aug 21, 2024

Primary Entity

Intuitive Surgical

Robotic Platform

da Vinci (and Ion)

Jurisdiction / Location

Hong Kong, China

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

FDA Class 2 recall Z-0964-2026 (firm-initiated Aug 21 2024; posted Dec 19 2025): Error 23062 due to partially damaged or failed surgeon console column motor connector. Causes ergonomics to lock — viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted.

System & Fleet Architecture

Operating Manufacturer

Intuitive SurgicalView Manufacturer

Hardware Platform

da Vinci (and Ion)

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=217489

    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=217489

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in Intuitive Surgical recalls 144 da Vinci 5 systems over surgeon console motor connector failure?

FDA Class 2 recall Z-0964-2026 (firm-initiated Aug 21 2024; posted Dec 19 2025): Error 23062 due to partially damaged or failed surgeon console column motor connector. Causes ergonomics to lock — viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. May lead to delay, aborted, or alternate procedure. FDA-determined cause: Device Design. 144 units in commerce across 30+ US states. Remedy: motor module sub-assembly replacement.

When did this incident occur?

The incident is recorded as occurring on August 21, 2024 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in Intuitive Surgical recalls 144 da Vinci 5 systems over surgeon console motor connector failure?

da Vinci (and Ion) by Intuitive Surgical is the recorded robot involved in this incident at Hong Kong, China.

Has anyone responded to Intuitive Surgical recalls 144 da Vinci 5 systems over surgeon console motor connector failure?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of Intuitive Surgical recalls 144 da Vinci 5 systems over surgeon console motor connector failure?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.