Intuitive Surgical recalls 12 da Vinci 5 systems over UI application crash and loss of monitor display

[RECALL]

Nov 17, 2025 · Intuitive Surgical · da Vinci (and Ion) · Hong Kong, China

Failure Mode

RECALL

Event Date

Nov 17, 2025

Primary Entity

Intuitive Surgical

Robotic Platform

da Vinci (and Ion)

Jurisdiction / Location

Hong Kong, China

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

FDA Class 2 recall Z-1428-2026 (firm-initiated Nov 17 2025; posted Feb 18 2026): Software version P1.2.1 caused an increase in Error 48306 events on da Vinci 5 systems in dual console configuration, causing User Interface application crashes and loss of interaction at the affected node. External monitor and tower monitor UI may not display (live endoscopic video still displayed).

System & Fleet Architecture

Operating Manufacturer

Intuitive SurgicalView Manufacturer

Hardware Platform

da Vinci (and Ion)

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=217883

    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=217883

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in Intuitive Surgical recalls 12 da Vinci 5 systems over UI application crash and loss of monitor display?

FDA Class 2 recall Z-1428-2026 (firm-initiated Nov 17 2025; posted Feb 18 2026): Software version P1.2.1 caused an increase in Error 48306 events on da Vinci 5 systems in dual console configuration, causing User Interface application crashes and loss of interaction at the affected node. External monitor and tower monitor UI may not display (live endoscopic video still displayed). FDA-determined cause: Software Design Change. 12 units in commerce across 6 US states. Corrective patch P1.2.2 in development.

When did this incident occur?

The incident is recorded as occurring on November 17, 2025 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in Intuitive Surgical recalls 12 da Vinci 5 systems over UI application crash and loss of monitor display?

da Vinci (and Ion) by Intuitive Surgical is the recorded robot involved in this incident at Hong Kong, China.

Has anyone responded to Intuitive Surgical recalls 12 da Vinci 5 systems over UI application crash and loss of monitor display?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of Intuitive Surgical recalls 12 da Vinci 5 systems over UI application crash and loss of monitor display?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.