Stryker Mako Total Knee software correction over 'error #3' codes when switching applications

[RECALL]

May 28, 2024 · Stryker · Stryker Mako SmartRobotics · Global

Failure Mode

RECALL

Event Date

May 28, 2024

Primary Entity

Stryker

Robotic Platform

Stryker Mako SmartRobotics

Jurisdiction / Location

Global

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

Voluntary software correction (announced ~May 28 2024; Class 2): increased 'software error #3' codes when switching between the total-knee and total-hip applications without a shutdown/restart, risking loss/deterioration of robotic function, surgical delay, or conversion to manual. Mitigation = shutdown/restart before switching apps. Cap-flag: exact units/Z-number not captured; other Mako accessory Class-2 recalls exist (handpiece, tracking kits, saw attachment) not individually verified here.

System & Fleet Architecture

Operating Manufacturer

Hardware Platform

Stryker Mako SmartRobotics

Operational Sectors

Physical AI / Robotics

Primary Sources & Evidentiary Filings

  • https://www.massdevice.com/stryker-warns-software-error-mako-surgical-robot/

    https://www.massdevice.com/stryker-warns-software-error-mako-surgical-robot/

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in Stryker Mako Total Knee software correction over 'error #3' codes when switching applications?

Voluntary software correction (announced ~May 28 2024; Class 2): increased 'software error #3' codes when switching between the total-knee and total-hip applications without a shutdown/restart, risking loss/deterioration of robotic function, surgical delay, or conversion to manual. Mitigation = shutdown/restart before switching apps. Cap-flag: exact units/Z-number not captured; other Mako accessory Class-2 recalls exist (handpiece, tracking kits, saw attachment) not individually verified here.

When did this incident occur?

The incident is recorded as occurring on May 28, 2024 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in Stryker Mako Total Knee software correction over 'error #3' codes when switching applications?

Stryker Mako SmartRobotics by Stryker is the recorded robot involved in this incident at Global.

Has anyone responded to Stryker Mako Total Knee software correction over 'error #3' codes when switching applications?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of Stryker Mako Total Knee software correction over 'error #3' codes when switching applications?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.