THINK Surgical recalls TMINI tracking arrays due to loosening causing potential misalignment

[RECALL]

May 8, 2026 · THINK Surgical · THINK Surgical TMINI · Fremont

Failure Mode

RECALL

Event Date

May 8, 2026

Primary Entity

THINK Surgical

Robotic Platform

THINK Surgical TMINI

Jurisdiction / Location

Fremont

Audit Standard

Physical State Verified

Incident Abstract & Audit Findings

THINK Surgical initiated a recall (Z-2716-2026) of TMINI Femur and Tibia Tracking Arrays (REF 109003/109004) because the knob on the tracking array may loosen, causing potential misalignment that could result in incorrect positional information communicated to the TMINI Miniature Robotic System. There have been 11 complaints regarding loose tracking arrays. Customers were instructed to inspect arrays before each use and return loose units.

System & Fleet Architecture

Operating Manufacturer

THINK SurgicalView Manufacturer

Hardware Platform

THINK Surgical TMINIView Spec

Operational Sectors

Primary Sources & Evidentiary Filings

  • https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=99219

    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=99219

Related Incidents & Failure Precedents

6 records on file

Incident Context & FAQ

What happened in THINK Surgical recalls TMINI tracking arrays due to loosening causing potential misalignment?

THINK Surgical initiated a recall (Z-2716-2026) of TMINI Femur and Tibia Tracking Arrays (REF 109003/109004) because the knob on the tracking array may loosen, causing potential misalignment that could result in incorrect positional information communicated to the TMINI Miniature Robotic System. There have been 11 complaints regarding loose tracking arrays. Customers were instructed to inspect arrays before each use and return loose units.

When did this incident occur?

The incident is recorded as occurring on May 8, 2026 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.

What robot was involved in THINK Surgical recalls TMINI tracking arrays due to loosening causing potential misalignment?

THINK Surgical TMINI by THINK Surgical is the recorded robot involved in this incident at Fremont.

Has anyone responded to THINK Surgical recalls TMINI tracking arrays due to loosening causing potential misalignment?

No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.

What is the current status of THINK Surgical recalls TMINI tracking arrays due to loosening causing potential misalignment?

This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.